Hioki 3030-10

Hioki 3030-10 HiTester Analog Multimeter User Manual

ແບບ ຈຳ ລອງ: 3030-10

1. ບົດແນະນຳ

This manual provides essential instructions for the safe and effective operation of the Hioki 3030-10 HiTester Analog Multimeter. This device is an average-sensing, manual-ranging meter designed for measuring voltage, resistance, and current in various electrical testing applications. Please read this manual thoroughly before use and retain it for future reference.

2. ຂໍ້ມູນຄວາມປອດໄພ

To ensure safe operation and prevent damage to the device, observe the following safety precautions:

3. ເນື້ອໃນຊຸດ

ກວດ​ສອບ​ວ່າ​ມີ​ລາຍ​ການ​ທັງ​ຫມົດ​:

4. ຜະລິດຕະພັນເກີນview ແລະຄຸນສົມບັດ

Hioki 3030-10 HiTester Analog Multimeter with test leads and carrying case

ຮູບທີ 1: Hioki 3030-10 HiTester Analog Multimeter. The image displays the multimeter unit, its protective carrying case, and the red and black test leads. The multimeter features a clear analog display with multiple scales for various measurements, a central rotary switch for function and range selection, and input jacks for the test leads.

The Hioki 3030-10 is an analog multimeter designed for reliable electrical measurements. Key features include:

5. ຕັ້ງຄ່າ

5.1 ການຕິດຕັ້ງຫມໍ້ໄຟ

The Hioki 3030-10 requires two AAA batteries for operation. To install or replace batteries:

  1. Ensure the multimeter is turned OFF (rotary switch set to 'OFF').
  2. ຊອກຫາຝາປິດຊ່ອງໃສ່ຫມໍ້ໄຟຢູ່ດ້ານຫຼັງຂອງໜ່ວຍ.
  3. Open the cover and insert two AAA batteries, observing the correct polarity (+/-) as indicated inside the compartment.
  4. ປິດຝາປິດຝາຫມໍ້ໄຟ.

5.2 ການເຊື່ອມຕໍ່ສາຍທົດສອບ

Connect the detachable test leads to the multimeter:

6. ຄໍາແນະນໍາການດໍາເນີນງານ

The Hioki 3030-10 is a manual-ranging multimeter. Always select the appropriate function and range using the rotary switch before making a measurement.

6.1 DC Voltage ການວັດແທກ (V)

  1. ຕັ້ງສະວິດ rotary ເປັນ DC Vol ທີ່ຕ້ອງການtage (V) range (e.g., 0.3V, 3V, 12V, 30V, 120V, 300V, 600V). Start with the highest range if the voltage ແມ່ນບໍ່ຮູ້ຈັກ.
  2. ເຊື່ອມຕໍ່ສາຍທົດສອບສີແດງເຂົ້າກັບດ້ານບວກ (+) ຂອງວົງຈອນ ແລະ ສາຍທົດສອບສີດຳເຂົ້າກັບດ້ານລົບ (-).
  3. ອ່ານ voltage value from the appropriate DCV scale on the analog display.

6.2 AC Voltage Measurement (~V)

  1. ຕັ້ງຄ່າສະຫຼັບ rotary ກັບ AC Vol ທີ່ຕ້ອງການtage (~V) range (e.g., 12V, 30V, 120V, 300V, 600V). Start with the highest range if the voltage ແມ່ນບໍ່ຮູ້ຈັກ.
  2. ເຊື່ອມຕໍ່ຕົວນໍາການທົດສອບໃນທົ່ວ AC voltage ແຫຼ່ງ.
  3. ອ່ານ voltage value from the appropriate ACV scale on the analog display.

6.3 DC Current Measurement (A)

ຂໍ້ຄວນລະວັງ: ເຊື່ອມຕໍ່ມັລຕິມີເຕີແບບອະນຸກົມກັບວົງຈອນສະເໝີເມື່ອວັດແທກກະແສໄຟຟ້າ. ຢ່າເຊື່ອມຕໍ່ມັນແບບຂະໜານກັນຂ້າມກະແສໄຟຟ້າtage ແຫຼ່ງ.

  1. Set the rotary switch to the desired DC Current (A) range (e.g., 60µA, 30mA, 300mA). Start with the highest range if the current is unknown.
  2. Break the circuit and connect the multimeter in series, ensuring the current flows through the meter.
  3. Read the current value from the appropriate DCA scale on the analog display.

6.4 ການວັດແທກຄວາມຕ້ານທານ (Ω)

ຂໍ້ຄວນລະວັງ: ໃຫ້ແນ່ໃຈວ່າວົງຈອນຖືກຕັດພະລັງງານກ່ອນທີ່ຈະວັດແທກຄວາມຕ້ານທານ.

  1. Set the rotary switch to the desired Resistance (Ω) range (e.g., x1, x10, x100, x1k).
  2. Before measuring, short the test leads together and adjust the "Ω ADJ" knob (if present) to zero the meter.
  3. ເຊື່ອມຕໍ່ຕົວນໍາການທົດສອບໃນທົ່ວອົງປະກອບທີ່ຈະວັດແທກ.
  4. Read the resistance value from the Ω scale on the analog display and multiply by the selected range multiplier.

6.5 Battery Test (BATT 1.5V)

  1. Set the rotary switch to the "BATT 1.5V" position.
  2. Connect the red test lead to the positive (+) terminal of the 1.5V battery and the black test lead to the negative (-) terminal.
  3. Read the battery charge level from the "BATT" scale on the analog display.

6.6 LED Test (LED)

  1. Set the rotary switch to the "LED" position.
  2. Connect the test leads to the LED. Observe polarity; if the LED does not light, reverse the leads.
  3. A lit LED indicates functionality.

6.7 ການວັດແທກອຸນຫະພູມ (TEMP)

This function requires a separately purchased thermistor probe. Refer to the thermistor probe's manual for specific connection and operation instructions.

  1. Set the rotary switch to the "TEMP" position.
  2. Connect the thermistor probe to the designated input jacks.
  3. Read the temperature from the "TEMP(°C)" scale on the analog display.

7. ບໍາລຸງຮັກສາ

7.1 ການທໍາຄວາມສະອາດ

ເຊັດແມັດຄasing ມີອ່ອນ, damp cloth. Do not use abrasive cleaners or solvents. Ensure the device is off and disconnected from all circuits before cleaning.

7.2 ການທົດແທນຟິວ

If the current measurement function ceases to operate, the fuse may need replacement. The Hioki 3030-10 uses a 0.5-amp/250V fuse. Always replace with a fuse of the identical type and rating.

  1. ໃຫ້ແນ່ໃຈວ່າມັນຕິມີເຕີປິດຢູ່ ແລະ ສາຍທົດສອບຖືກຖອດອອກແລ້ວ.
  2. Refer to the device's internal diagram (if available) or contact customer support for instructions on accessing the fuse compartment.
  3. Carefully remove the old fuse and insert the new 0.5A/250V fuse.
  4. Reassemble the meter securely.

7.3 ການເກັບຮັກສາ

When not in use, store the multimeter in its carrying case in a cool, dry place, away from direct sunlight and extreme temperatures. Remove batteries if storing for extended periods to prevent leakage.

8. ການແກ້ໄຂບັນຫາ

If the multimeter does not function as expected, consider the following:

If problems persist after attempting these steps, contact Hioki customer support for further assistance.

9. ຂໍ້ມູນຈໍາເພາະ

ພາລາມິເຕີມູນຄ່າ
ໝາຍເລກຕົວແບບ3030-10
ປະເພດການວັດແທກAnalog Multimeter (Average-Sensing, Manual-Ranging)
ຊິງຕັນດີຊີ Voltage ຊ່ວງ0.3V, 3V, 12V, 30V, 120V, 300V, 600V
AC Voltage ຊ່ວງ12V, 30V, 120V, 300V, 600V
DC Current Ranges60µA, 30mA, 300mA
ຊ່ວງການຕໍ່ຕ້ານx1, x10, x100, x1k (Max 3MΩ)
ການທົດສອບຫມໍ້ໄຟຫມໍ້ໄຟ 1.5V
ການປ້ອງກັນຟິວ0.5A / 250V
ມາດຕະຖານຄວາມປອດໄພIEC 61010, CAT III 600V
ແຫຼ່ງພະລັງງານ2 x ຫມໍ້ໄຟ AAA (ບໍ່ລວມ)
ນ້ໍາຫນັກລາຍການ150 g
ຂະໜາດສິນຄ້າ (L x W x H)15.75 x 15.49 x 5.84 cm
ການດໍາເນີນງານ Voltage (Min.)0.3 ໂວນ (DC)
ລະດັບອຸນຫະພູມສູງສຸດ150 ອົງສາ Celsius

ເອກະສານທີ່ກ່ຽວຂ້ອງ - 3030-10

ກ່ອນview HIOKI Digital HiTester 3237, 3238, 3239: High-Speed DMMs
Explore the HIOKI Digital HiTester series, featuring the 3237 (Economical), 3238 (Advanced), and 3239 (4-terminal Ω) models. These high-speed Digital Multimeters (DMMs) offer 3.3 ms sampling, comparator functions, and versatile interfaces for efficient production line testing and laboratory measurements.
ກ່ອນview HIOKI 3169-20/21 Clamp ຄູ່ມືເລີ່ມຕົ້ນດ່ວນ HiTester ເມື່ອເປີດເຄື່ອງ
ຄູ່ມືເລີ່ມຕົ້ນດ່ວນສຳລັບ HIOKI 3169-20/21 Clamp ໃນ Power HiTester, ເຊິ່ງກວມເອົາການຕັ້ງຄ່າ, ການລະບຸຊິ້ນສ່ວນຕ່າງໆ, ຂັ້ນຕອນການວັດແທກ, ແລະບັນທຶກຄວາມປອດໄພ. ໃຫ້ຄໍາແນະນໍາລະອຽດສໍາລັບການເຊື່ອມຕໍ່, ການຕັ້ງຄ່າ, ແລະການປະຕິບັດການວັດແທກພະລັງງານ.
ກ່ອນview ຄູ່ມືການໃຊ້ງານເຄື່ອງທົດສອບກະແສໄຟຟ້າຮົ່ວ HIOKI ST5540/ST5541
ຄູ່ມືການໃຊ້ງານນີ້ລະບຸລາຍລະອຽດກ່ຽວກັບການດຳເນີນງານ, ການຕັ້ງຄ່າ, ຂັ້ນຕອນການວັດແທກ ແລະ ແນວທາງຄວາມປອດໄພສຳລັບເຄື່ອງທົດສອບກະແສໄຟຟ້າຮົ່ວ HIOKI ST5540 ແລະ ST5541, ເຊິ່ງເປັນເຄື່ອງມືທີ່ຈຳເປັນສຳລັບການທົດສອບຄວາມປອດໄພທາງໄຟຟ້າ.
ກ່ອນview HIOKI 3169-20/21 Clamp On Power HiTester Instruction Manual
Comprehensive instruction manual for the HIOKI 3169-20 and 3169-21 Clamp On Power HiTesters. Covers product overview, features, setup, measurement procedures, specifications, and maintenance for electrical power analysis.
ກ່ອນview Hioki 3665-20 LAN Cable HiTester Instruction Manual
Comprehensive instruction manual for the Hioki 3665-20 LAN Cable HiTester. Learn how to test wiremaps, measure cable length, and identify network cable connections with this handheld device.
ກ່ອນview HIOKI 3174/3174-01 AC Automatic Insulation/Withstanding HiTester Instruction Manual
This instruction manual provides comprehensive guidance for the HIOKI 3174 and 3174-01 AC Automatic Insulation/Withstanding HiTesters. Learn about product features, safety precautions, testing procedures for insulation resistance and withstand voltage, external interface control, and maintenance.